Performance Tests
7
Agilent U1602A/U1604A User’s and Service Guide
181
Resistance Verification Test (2-Wire Ohms)
Calibrator Settings
Range
Test Limit
Results
600
W
600
W
595.8
W
to 604.2
W
6 k
W
6 k
W
5.958 k
W
to 6.042 k
W
60 k
W
60 k
W
59.58
k
W
to 60.42 k
W
600 k
W
600 k
W
595.8 k
W
to 604.2 k
W
3 M
W
6 M
W
2.973 M
W
to 3.027 M
W
6 M
W
6 M
W
5.958 M
W
to 6.042 M
W
60 M
W
60 M
W
59.28 M
W
to 60.72 M
W
Capacitance Verification Test
Calibrator Settings
Range
Test Limit
Results
60 nF
60 nF
58.68 nF to 61.32 nF
600 nF
600 nF
586.8 nF to 613.2 nF
6
m
F
6
m
F
5.868
m
F to 6.132
m
F
36
m
F
60
m
F
35.16
m
F to 36.84
m
F
60
m
F
60
m
F
58.68
m
F to 61.32
m
F
300
m
F
300
m
F
293.4
m
F
to 306.6
m
F
Diode Verification Test
Calibrator Voltage
Test Limit
Results
0.9 V
0.8812 V to 0.9188 V
Continuity Verification Test
Calibrator Settings
Condition
Results
30
W
Buzzer sounds on
[PASS/FAIL]
75
W
Buzzer sounds off
[PASS/FAIL]