General Disassemble
8
Agilent U1602A/U1604A User’s and Service Guide
193
To dismantle system boards
1
Perform the following procedures:
•
•
•
To remove handler
•
To remove front and rear casing
2
The front view of the PCBA board is shown in Figure 8- 10. Turn over
the PCB assembly board with the LCD display facing downward and
remove the screw at the location as pointed in Figure 8- 13.
Figure 8-13
Remove the six screws and pull the ribbon cable from the PCBA board
3
Gently disconnect the ribbon cable of the LCD display from the PCBA
board.
4
Unplug the pin connectors at both side of the PCBA board as pointed
in Figure 8- 13.
5
Gently segregate both the PCBA boards from each other.
•
The first PCBA board with LCD display unit attached is the multimeter
measurement board.
•
The second PCBA board is the scope measurement board.
Ribbon Cable
Pin Connectors