Performance Tests
7
Agilent U1602A/U1604A User’s and Service Guide
171
Resistance Verification Test
1
Put the calibrator in Standby (STBY) mode.
2
Select Meter mode of the instrument, then press F2 softkey for
Ohmmeter.
3
Press F1 to select resistance measurement.
4
Connect the calibrator to the instrument’s input and ground terminal as
shown in Figure 7- 2 using a banana plug.
5
Select each function and range in the order shown in Table 7- 7. Apply
the values from the calibrator shown in Table 7- 7 below.
6
Compare the measurement results to the corresponding test limits
shown in Table 7- 7.
Table 7-7
Resistance Verification Test
Calibrator Settings
Function
Range
Min Limit
Max Limit
600
Ω
2-Wire Ohms
600
Ω
595.8
Ω
604.2
Ω
6 k
Ω
6 k
Ω
5.958 k
Ω
6.042 k
Ω
60 k
Ω
60 k
Ω
59.58 k
Ω
60.42 k
Ω
600 k
Ω
600 k
Ω
595.8 k
Ω
604.2 k
Ω
3 M
Ω
6 M
Ω
2.973 M
Ω
3.027 M
Ω
6 M
Ω
6 M
Ω
5.958 M
Ω
6.042 M
Ω
60 M
Ω
60 M
Ω
59.28 M
Ω
60.72 M
Ω