Performance Tests
7
Agilent U1602A/U1604A User’s and Service Guide
177
Table 7-10
Self-calibration error code
Offset Error
Bit number
Description
TE4
[14]
Trigger level calibration error at the second input source for channel 2 in pattern trigger.
TE3
[13]
Trigger level calibration error at the second input source for channel 1 in pattern trigger.
TE2
[12]
Trigger level calibration error for trigger source at channel 2 in DC trigger coupling.
TE1
[11]
Trigger level calibration error for AC trigger coupling for both channels.
TE0
[10]
Trigger level calibration error for trigger source at channel 1 in DC trigger coupling.
OE9
[9]
Offset gain calibration error for channel 2
OE8
[8]
Offset gain calibration error for channel 1
OE7
[7]
Second zero offset calibration error for channel 2 at the range of 5 mV/div to 20 mV/div.
OE6
[6]
Second zero offset calibration error for channel 1 at the range of 5 mV/div to 20 mV/div.
OE5
[5]
Second zero offset calibration error for channel 2 at the range of 50 mV to 100 V/div.
OE4
[4]
Second zero offset calibration error for channel 1 at the range 50 mV to 100 V/div.
OE3
[3]
First zero offset calibration error for channel 2 at 125 ns/div
OE2
[2]
First zero offset calibration error for channel 1 at 125 ns/div
OE1
[1]
First zero offset calibration error for channel 2 at the following conditions:
•
all time base divisions except 125 ns/div
•
when using both channels at 125 ns/div
OE0
[0]
First zero offset calibration error for channel 1 at the following conditions:
•
all time base divisions except 125 ns/div
•
when using both channels at 125 ns/div
N O T E
•
OE — Zero offset calibration error
•
TE — Trigger level calibration error