6
Service and Maintenance
154
Agilent U1602A/U1604A User’s and Service Guide
• If there is no trace display
✔
Recall factory default setting by pressing
button on the front
panel. Press F1 to enter Save/Load Setup menu, in the page 1/4, then
press F1 to restore factory default setting.
✔
Check that the scope probe lead wires are securely inserted into the
connector assembly and that the probe clips are making good contact
with the probe lead wires.
✔
Check that the circuit under test is powered on, probe clips are
securely connected to test points on the circuit under test, and ground
is connected.
✔
Press Autoscale button to allow the instrument to automatically trigger
the signal to the best display.
✔
Contact Agilent Technologies Service Center to obtain technical
assistance, if necessary.
• If the trace display is unusual or unexpected
✔
Check that the horizontal time/division is correctly set for the expected
frequency range of the input signals.
✔
Check that all scope probes are connected to the correct signals on the
circuit under test, and ground lead is securely connected to ground of
the circuit under test.
✔
The trigger setup is the most important factor in capturing the signal
you desire. Check that the trigger setup is correct.
✔
Check that infinite persistence in the Display menu is turned off. Press
Clear Display to remove the display.
✔
Press Autoscale button to allow the instrument to automatically trigger
the signal to the best display.
✔
Contact Agilent Technologies Service Center to obtain technical
assistance, if necessary.