BER Frames
TESTs:MANual:SENSitivity:FRAMe <int>
TESTs:MANual:SENSitivity:FRAMe?
This command sets the number of the frames for bit error ratio
(BER) tests and can be used in Stand-by or Connection status.
The allowable range is from
1
to
4100
(frames) in
1
step. At
preset, this is set to
10
.
This query command returns the set number of the frames for
BER tests with integer.
OCNS
TESTs:MANual:OCNS ON|OFF|1|0
TESTs:MANual:OCNS?
This command sets whether or not to enable OCNS. The
parameter is
ON
(1) or
OFF
(0). This is valid in the Connection
(RMC) status. Also, this is reset to
OFF
(0) when the status
becomes Stand-by.
This query command returns the setting of OCNS function with
the value
1
or
0
. The meaning of the value is same as that of the
command parameter.
Caller ID
TESTs:MANual:CALLer:ID <string>
TESTs:MANual:CALLer:ID?
This command sets the Caller ID on the [Stand-by] screen.
<string>: Sets dialing number at BS call(AMR) test step. Up to
20 digit arbitrary number (0 to 9) and symbol * or #. This is not
initialized by PRESET execution with the Starting Condition
softkey.
This query command returns the set Caller ID with string.
Averaging
TESTs:MANual:AVERage OFF|<int>
TESTs:MANual:AVERage?
Содержание N9360A
Страница 1: ...Agilent Technologies Agilent N9360A Multi UE Tester W CDMA Programming Manual...
Страница 4: ...1 4 N9360A Multi UE Tester W CDMA Programming Manual 1 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 8: ...8 N9360A Multi UE Tester W CDMA Programming Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 12: ...6 N9360A Multi UE Tester W CDMA Programming Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...1 4 N9360A Multi UE Tester W CDMA Programming Manual 1 Legal Information THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 180: ...8 20 N9360A Multi UE Tester W CDMA Programming Manual 8 Syntax Diagrams THIS PAGE IS INTENTIONALLY LEFT BLANK...