5-20
N9360A Multi UE Tester W-CDMA Programming Manual
5
Programming Command Reference
These query commands return the set Band and channel
number with integer. The range of the value is same as that of
the command.
Open Loop TX Power
CONFigure:SEQuence:OPENpower <int>
CONFigure:SEQuence:OPENpower?
This command sets whether or not to include Open Loop TX
Power test in the test sequence.
<int>
is
1
for Run or
0
for —
(skip). At preset, this is set to
1
.
Open Loop TX Power test can be executed only on
RFCH1
.
This query command returns the setting of the Open Loop TX
Power test with integer. The meaning of the return value is same
as that of the command parameter.
Inner Loop Power
CONFigure:SEQuence:INNErpower <int>
CONFigure:SEQuence:INNErpower?
This command defines whether or not to include Inner Loop
Power test in the test sequence.
<int>
is a sum of the values of
all the 6 bits as follows.
bit 0 (1): For testing RFCH1
bit 1 (2): For testing RFCH2
bit 2 (4): For testing RFCH3
bit 3 (8): For testing RFCH4
bit 4 (16): For testing RFCH5
Table 5-1
RFCH Allowable Range and Preset Value
Int
Allowable Range
Preset Value
CH1
CH2
CH3
CH4
CH5
CH6
1
–1
1
6
1
1
1
—
—
—
2
99999
9600 to 9900
812|837|4150 to 4200
9612
9750
9888
—
—
—
Содержание N9360A
Страница 1: ...Agilent Technologies Agilent N9360A Multi UE Tester W CDMA Programming Manual...
Страница 4: ...1 4 N9360A Multi UE Tester W CDMA Programming Manual 1 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 8: ...8 N9360A Multi UE Tester W CDMA Programming Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 12: ...6 N9360A Multi UE Tester W CDMA Programming Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...1 4 N9360A Multi UE Tester W CDMA Programming Manual 1 Legal Information THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 180: ...8 20 N9360A Multi UE Tester W CDMA Programming Manual 8 Syntax Diagrams THIS PAGE IS INTENTIONALLY LEFT BLANK...