How to Measure Power
Power Measurement
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
83
Figure 52 shows the possible range values you can choose. These values range
from 10 dBm (upper power limit of 13 dBm) to
−
70 dBm (upper power limit of
−
67 dBm) in 10 dBm increments.
Upper Power Limit and Resolution
Table 1 shows the upper power limit and measurent resolution at various power
ranges. As can be seen the resolution decreases as the chosen [
Range
] decreases.
The resolution is always 40 dBm less than the chosen [
Range
] value. The Upper
Power Limit is always 3 dBm greater than the chosen [
Range
] value.
Figure 52
Range Value Menu
Table 1
Upper Power Limits and Resolution for Various Power Ranges
[R
ANGE
]
Upper Power Limit
Resolution
10 dBm
19.999 mW
13.000 dBm
0.001 mW
−
30.000 dBm
0 dBm
1.9999 mW
3.000 dBm
0.1
µ
W
−
40.000 dBm
−
10 dBm
199.99
µ
W
−
7.000 dBm
0.01
µ
W
−
50.000 dBm
−
20 dBm
19.999
µ
W
−
17.000 dBm
0.001
µ
W
−
60.000 dBm
−
30 dBm
1.9999
µ
W
−
27.000 dBm
0.1 nw
−
70.000 dBm
−
40 dBm
199.99 nW
−
37.000 dBm
0.01 nw
−
80.000 dBm
−
50 dBm
19.999 nW
−
47.000 dBm
0.001 nw
−
90.000 dBm
−
60 dBm
1.9999 nW
−
57.000 dBm
0.1 pW
−
100.000 dBm
−
70 dBm
199.99 pW
−
67.000 dBm
0.01 pW
−
110.000 dBm