How to Use Triggers
Tunable Lasers
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
127
2
Move to the Tunable Laser channel and press [Menu].
3
Move to <
Output Trigger Mode
>, by using the cursor key, and press
Enter
.
You will see the screen in Figure 82.
4
Move to one of the following, by using the cursor key:
–
<Disabled
>, the output trigger mode is disabled.
–
<
Modulation
>, the output trigger connector outputs a TTL signal at the
frequency of the internal modulation. This signal is output whether the laser
is switched on or off.
–
<
Step Finished
>, a trigger is output after every step of a sweep finishes.
–
<
Sweep Finished
>, a trigger is output after a sweep cycle finishes.
–
<
Sweep Started
>, a trigger is output after a sweep cycle starts.
5
Press
Enter.
N O T E
If you choose <
Step Finished
> and a <
Continuous
> sweep, the wavelength
interval between hardware triggers is set by the [
Step
] parameter, although, the
sweep is not stepped.
Figure 82
Output Trigger Mode