Firmware Upgrades
Firmware Upgrade Process
292
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
6
Select the com port where the serial cable is connected. Choose the maximum
baud rate. Click on OK.
7
Turn on the 8166A and wait for the first beep.
Now you must be quick:
–
Press and hold the
Preset
hardkey.
–
Still holding
Preset
, press and hold the
Aux
hardkey.
–
Still holding
Preset
and
Aux
, press and hold the
Appl
hardkey.
T I P
Use three fingers and place them near the keys in advance!
If you were successful, you here a double beep and the display shows that the
instrument is in download mode. Otherwise, turn off the power and try again.
8
Click on ’Program’ and wait until the download completes. The instrument will
reboot with the new firmware.