The PACT Application
Applications
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
183
How to Set Up PACT
To set up PACT:
1
Insert a Tunable Laser module and at least one Power Meter module into your
Agilent 8163A Lightwave Multimeter or Agilent 8164A Lightwave
Measurement System.
2
Turn your instrument on.
3
Press the
Appl
hardkey. The Applications menu, as shown in Figure 117,
appears.
4
Move to <
PACT
> and press
Enter
. The PACT Setup Screen, as shown in
Figure 135, appears.
Selecting a Tunable Laser Module
5
Move to the Module Selection box. As a default, the Tunable Laser module
with the lowest slot number is selected. Move to [
Laser Selection
]. Press
[Edit].
6
To select a Tunable Laser module, press [Edit], use the cursor key to move to
your preferred Tunable Laser module as denoted by slot and channel numbers,
see
“Slot and Channel Numbers” on page 36
, and press
Enter
.
7
Perform steps until you have selected the required Tunable Laser modules for
your application and press
Enter
.
N O T E
If no Tunable Laser module is selected, you cannot exit by pressing
Enter
.
Figure 135
The PACT Setup Screen