![Agilent Technologies Agilent 8163A Скачать руководство пользователя страница 109](http://html.mh-extra.com/html/agilent-technologies/agilent-8163a/agilent-8163a_user-manual_2867997109.webp)
How to Set the Power
Tunable Lasers
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
109
– <
Off
>, to disable the dark position
4 Press
Enter
to return tothe Menu screen.
5 Press [Close] to exit from the Menu screen.
The Analog Output
If there is an output BNC connector on the front panel of your Tunable Laser
module, you can output a DC voltage level that is proportional to the laser output
power. The relationship between this voltage level and the output power is not
calibrated.
Attenuator
If your Tunable Laser module has an attenuator installed, the Analog
Output signal depends on the power mode you have chosen. The voltage level
is proportional to the optical power produced by the laser diode (not the
optical output power).
In <
Manual Att.
> Mode, the relationship between the power value, [P], and
the voltage level is only proportional to the output power if the attenuation
remains constant. The attenuator adds a constant offset to the voltage level.
In <
Automatic
> Attenuation Mode, the laser output power, and therefore
the Analog Output signal, is not linearly proportional to the output power.
This is because the actual output optical power is derived by the optimum
combination of the laser output power and the attenuation.
To set the power to
−
10 dBm:
1
Move to power parameter, [P], and press [Pwr unit]. You should see the power
unit menu.
2
Move to <dBm> and press
Enter
.
3
Press
Enter
to edit the power value.
4
Enter
−
10.000 and press
Enter
.
Figure 69
Output Power and the Analog Output in <M
ANUAL
A
TT
.> Mode
Voltage, V
[P], mW
5 V
20 mW
2 mW
[A
TTEN
.] = 0 dB
[A
TTEN