Getting Started
How to Change the Value of a Parameter
48
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
5
When you have finished editing the value, press
Enter
. The edited value
becomes the new value of the parameter.
To change the wavelength from 1540.000 nm to 1525.000 nm:
1
Move to the wavelength parameter for a Tunable Laser module and press
[Edit]. The most significant digit is highlighted.
2
Press the left cursor once to highlight the digit four.
3
Press the down cursor twice to change the value of the digit to two.
4
Press the right cursor once to move the cursor one digit right.
5
Press the up cursor five times to change the value of the digit to five. Press
Enter
to end the editing.
How to Change a Discrete Parameter
For discrete parameters, you may choose a particular values within a given range.
For a Power Sensor module:
1
Move to the Power Sensor channel and press the [Details] softkey.
2
Move to the [
AvgTime
] parameter and press
Enter
.
3
Move to 1 s, by using the cursor key, and press
Enter
.
or
1
Move to the Power Sensor channel and press the [Menu] softkey.
2
Move to the <
Averaging Time
> parameter and press
Enter
. You see the screen
in Figure 26.
3
Move to <1 s>, by using the cursor key, and press
Enter
.