Additional Information
Using the System Utilities
62
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
If You Forget Your Password
If you forget your password, contact your nearest Agilent Technologies
Sales/Service Office. Find contact details by accessing http://www.agilent.com/
How to Set the Trigger Configuration
The trigger level at the external trigger connectors is by default active high, this
means when a trigger rises above the high TTL level, a trigger is accepted.
You can select three modes of triggering from the trigger configuration menu:
•
<
None
>, which you should choose if you do not want to use triggering.
•
<
Default
>, which you should choose if you want to enable the trigger
connectors.
•
<
Pass Through
>, which you should choose if you want an input trigger to
automatically generate an output trigger. This allows you to trigger another
instrument almost simultaneously.
•
<
Loopback
>, which you should choose if you want an output trigger to
automatically generate an input trigger. For example, using this mode, you
could trigger each step of a wavelength sweep with just one externally
generated input trigger.
To change the triggering mode:
1
Press the
Config
hardkey.
2
Move to the <
Trigger
> menu option and press
Enter
. You see a box displaying
the available triggering modes.
Figure 36
Changing the Triggering Mode