![Agilent Technologies Agilent 8163A Скачать руководство пользователя страница 8](http://html.mh-extra.com/html/agilent-technologies/agilent-8163a/agilent-8163a_user-manual_2867997008.webp)
8
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
Initial Safety Information for Tunable Laser Modules
Laser Safety Labels - 28 mW Max. CW Output Power
Table 3
Tunable Laser Modules Laser Safety Information
Agilent 81680A Agilent 81682A Agilent 81640A
Agilent 81689A
Laser Type
FP-Laser
InGaAsP
FP-Laser
InGaAsP
FP-Laser
InGaAsP
FP-Laser
InGaAsP
Wavelength range
1400-1670 nm
1400-1670 nm
1400-1670 nm
1400-1670 nm
Max. CW output power*
<15 mW
<15 mW
<15 mW
<15 mW
Beam waist diameter
9
µ
m
9
µ
m
9
µ
m
9
µ
m
Numerical aperture
0.1
0.1
0.1
0.1
Laser Class according to
IEC 60825-1 (1998)- International
3A
3A
3A
3A
Max. permissible CW output power - IEC**
50 mW
50 mW
50 mW
50 mW
Laser Class according to
FDA 21 CFR 1040.10 (1995) - USA
IIIb
IIIb
IIIb
IIIb
Max. permissible CW output power - FDA**
500 mW
500 mW
500 mW
500 mW
*
Max. CW output power
is defined as the highest possible optical power that the laser source can produce at its output
connector.
** Max. permissible CW output power
is the highest optical power that is permitted within the appropriate laser class.
Figure 1
USA Safety Labels (81655A, 81656A, 81657A)