How to Perform a Wavelength Sweep
Tunable Lasers
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
115
6
Move to
[λ
Stop
] and press
Enter
.
7
Enter 1570.000, press
Enter
.
8
Move to [
Cycles
] and press
Enter
.
9
Enter 3 and press
Enter
.
10
Move to [
Repeat Mode
] and press
Enter
.
11
Move to <
Twoway
>, by using the cursor key, and press
Enter
.
12
Move to
[
Step
] and press
Enter
.
13
Enter 1.000 and press
Enter
.
14
Move to
[
Dwell
] and press
Enter
.
15
Enter 0.5000 and press
Enter
.
16
Press the [Active] hardkey on the front panel of your Tunable Laser module to
enable the optical output.
17
Press [Run Swp] to start the sweep. The screen in Figure 74 appears.
18
The wavelength is swept automatically but you can:
a
press [Stop] to end the sweep, or
b
press [Pause] to pause the sweep. The screen in Figure 75 appears. You
can:
–
Press [Continue] to continue sweeping automatically.
–
Press [Prev] or [Next] to sweep manually.
–
Press [Stop] to end the sweep.