The PACT Application
Applications
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
187
20
Press [New Ref]. A box similar to the box in Figure 137 appears prompting
you to connect a fiber to the first Power Meter channel.
21
Connect the Tunable Laser module to the Power Meter channel through all the
system components, but excluding the Device Under Test. Press [Next] to start
the reference measurement.
The PACT Reference Measurement screen, as shown in Figure 138, appears.
A progress bar at the bottom of the screen shows the proportion of the stability
application that has been completed. The wavelength,
λ
, shows you the current
wavelength of the Tunable Laser module.
22
There are two softkeys available from the PACT Reference Measurement
screen.
–
Press [Graph] to view a graph of the Measurement results as the results are
measured. Press [Close] to return to the PACT Reference Measurement
screen. See
“Working with Application Graphs” on page 155
information on zooming in and out.
–
Press [Cancel] to return to the PACT Reference screen without completing
the reference measurement, all reference values that were measured will be
deleted.
Figure 137
Connection Prompt Screen