![Agilent Technologies 969-8888 Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/agilent-technologies/969-8888/969-8888_instruction-manual_2868751088.webp)
TECHNICAL INFORMATION
82/107
87-901-011-01
Foreline flange
KF 40 NW
Nominal rotational
speed
31800 rpm
Start-up time without
gas load and with the
recommended fore-
pump
<
6 minutes
Minimum recom-
mended forepump
At least 60 m
3
/h (but depending on
the inlet gas flow)
Operating position
Any
Operating ambient
temperature
+5 °C to +35 °C
Max rotor temperature
120 °C
Vibration level
(displacement)
<
0.01
μ
m at inlet flange
Lubricant permanent
lubrication
Cooling requirements
Water
Coolant water
Recommended flow: 200 l/h
Temperature: +15 °C to +30 °C
Pressure: 3 to 5 bar (45 to 75 psi)
Noise level
<60 dB(A) at 1 meter
Power supply:
Input
voltage:
Input
freq.:
Max input power:
Stand-by
power:
Max
operating
power:
208 - 240 Vac
50 - 60 Hz
950 VA
30 to 35 W
515 W for Nitrogen and lighter
gases
430 W for Argon with purge on
390 W for Argon with purge off
Protection fuse
2 x 12,5 A
Compliance with:
•
EN 55011 (Radio disturbance
characteristics)
•
EN 61000-3-2 (Limits for har-
monic current emissions)
•
EN 61000-3-3 (Limitation of volt-
age changes, voltage fluctua-
tions and flicker)
•
EN 61000-6-2 and EN 61326
(Immunity for industrial envi-
ronments) which recall:
- EN 61000-4-2 (Electrostatic
discharge phenomena)
- EN 61000-4-3 (Radiated
high-frequency phenomena)
- EN 61000-4-4, EN 61000-4-
5, EN61000-4-6 (Conducted
high-frequency phenomena)
- EN 61000-4-11 (Voltage
fluctuation, voltage dips and
short interruptions)
Installation Category
II
Pollution Degree
2
Power cable
With European or NEMA plug
3 meters long (optional)
Serial communication
kit
RS232 cable with a 9-pin D type
male connector and a 9-pin D type
female connector, and T-Plus soft-
ware (optional)
Storage temperature
-20° C to +70° C
Environment protection
IP54
Weight (ISO 250)
kg
(lbs)
52 (114.4)
NOTE
When the Turbo-V 3K-G has been stored at a tem-
perature less than 5°C, wait until the system has
reached the above mentioned ambient temperatures
before starting the pump.