TECHNICAL INFORMATION
99/107
87-901-011-01
between the ProfiBus network and the pump’s
controller serial interface. As a consequence of
this, the scale and the meaning of some parame-
ters depends directly to the specific model of
Turbo Pump controller connected to the interface.
PROFIBUS MESSAGE MAPPING
PARAMETER BUFFER (17 BYTES)
OFFSET
BYTE
SIZE
BYTE
UNIT
RES
RANGE DESCRIPTION
WIN
SERIAL
0
1
°C
15-35
Water Cooling Max Temperature
(not used – insert 20)
158
1
1
°C
1-5
Water Cooling Hysteresis
(not used – insert 2)
159
2 1 %
1-99
Setpoint
Hysteresis
Set the hysteresis on set point output (see set point out paragraph)
105
3
1
°C
-
Heather Jacket Temperature Setting
Specific option, not available on standard models
113
4 1
sccm
-
Flowmeter
Threshold
Specific option, not available on standard models
115
5 2
RESERVED
-
7 2
Hz/mA/s
-
Setpoint
Value
Set the threshold for the set point out
102
9
2
s
0-99999 Setpoint Mask Time
Set the time during the controller does not drive the set point out
103
11 2
0.2*s
0-65535
0 = inf.
Vent Valve Opening Delay
Set the delay time between the stop command and vent opening.
Valid only in “purge/vent auto mode” (Parameter Buffer byte-15 B1=0); de-
fault = 15; 0 = infinite (the valve will never open)
126
13 2
0.2*s
0-65535
0=inf
Vent Valve Open Time
Set how long the vent will remain open, then both purge and vent will be
closed. Valid only in “purge/vent auto mode” (Parameter Buffer byte-15
B1=0); 0 = infinite (the valve remain open till a start command)
147
B0 - Vent & Purge Controlled Vent Procedure/Man
0 = Controlled Vent Procedure
1 = Manual (see Par. Buff. byte-15-B1)
125
B1 - Vent & Purge Remote/Profibus
0 = serial or Profibus (see Out. Buff. byte-0 B3 and B2)
1 = remote ( the valves are driven by the inputs on remote I/O connector)
146
B2 - Interlock Type
0 = continuos (the interlock must be present to run the pump)
1 = at start (the interlock must be present only at start command)
110
B3 - Setpoint Logic
0 = active close (NO)
1 = active open (NC)
104
B4 - Heather Jacket Mode (Auto/man)
Specific option, not available on standard models
-
B5 - Gas Load Type
0 = Ar
1 = N
2
157
B6 - Purge Gas Type
0 = N
2
1 = Ar
-
15 1
Bit
Field
B7 - Reserved
114