Installation
Agilent
8169A
Polarization Controller, Second Edition
101
Replacing the Fuse
There is one fuse in this instrument. This is a T1A/250V (time-lag) (Agilent
Part No. 2110-0007). The fuse holder is at the rear of the instrument,
beside the line power connector. To replace the fuse,
1
Release the fuse holder: use the blade of a at-headed screwdriver to depress the
catch at the side of the holder and then pull the holder out a little.
Figure 19 Releasing the Fuse Holder
2
Pull the fuse holder out of the instrument.
Figure 20 The Fuse Holder
3
Check and replace the fuse as necessary making sure that the fuse is always in
the top position of the fuse holder, and the bridge is in the bottom.
4
Place the fuse holder back in the instrument, and push it until the catch clicks
back into place.