Setting the State of Polarization
Remote Commands
Agilent
8169A
Polarization Controller, Second Edition
71
[:INPut]:POSition:QUARter
Syntax
[:INPut]:POSition:QUARter <wsp>
<value>jMINimumjMAXimumjDEFault where value is a oating point
number between -360.00 and 360.00.
Description
This command sets the position of the
λ
/4 retarder plate. The parameter
may be either
• a number, in mechanical degrees (do not give a unit; the number will be rounded
to the nearest 0.05°),
• MINimum (-360.00°),
• MAXimum (360.00°), or
• DEFault (0.00°).
[:INPut]:POSition:QUARter?
Syntax
[:INPut]:POSition:QUARter?
Description
This query gets the position of the
λ
/4 retarder plate in mechanical degrees
(without a unit).
Example
OUTPUT 724;":POS:QUAR 64"
OUTPUT 724;":POS:HALF 99.5"
OUTPUT 724;":POS:QUAR?"
ENTER 724;Q$
OUTPUT 724;":POS:HALF?"
ENTER 724;H$
Содержание 8169A
Страница 36: ...Scanning the Poincare Sphere Analyzing the Results 38 Agilent 8169A Polarization Controller Second Edition ...
Страница 46: ...Programming the Polarization Controller GPIB Interface 48 Agilent 8169A Polarization Controller Second Edition ...
Страница 82: ...Remote Commands SYSTem Commands 84 Agilent 8169A Polarization Controller Second Edition ...
Страница 106: ...Installation 108 Agilent 8169A Polarization Controller Second Edition ...
Страница 114: ...Specifications 116 Agilent 8169A Polarization Controller Second Edition Declaration of Conformity ...
Страница 158: ...Cleaning Information Other Cleaning Hints 160 Agilent 8169A Polarization Controller Second Edition ...
Страница 166: ...Error Messages 168 Agilent 8169A Polarization Controller Second Edition ...
Страница 167: ......