Setting a state of Polarization
Setting the Optimum Transmission SoP
28
Agilent
8169A
Polarization Controller, Second Edition
Setting the Optimum
Transmission SoP
1
Read the value for
ε
B
from the display.
2
Add 180° to this value.
3
Type in the new value, and press
The state of polarization is now set to the value for the current setup that
gives the greatest power through the fiber. This is possible here because
the fiber behaves linearly. For non-linear components the polarizations for
worst case and optimum transmission will not be on opposite sides of the
sphere, and the angle between them is a characteristic of the component.