Setting Up the STATus Registers
Remote Commands
Agilent
8169A
Polarization Controller, Second Edition
77
Description
This command sets the bits in the ENABle register that enable the
contents of the EVENt register to affect the Status Byte (STB). Setting a bit
in this register to 1 enables the corresponding bit in the EVENt register to
affect bit 7 of the Status Byte.
:STATus:OPERation:ENABle?.
Syntax
:STATus:OPERation:ENABle?
Description
The query returns the current contents of the OPERation:ENABle register.
Example
OUTPUT 724;”STAT:OPER:NTR 2”
OUTPUT 724;”STAT:OPER:PTR 256”
OUTPUT 724;”STAT:OPER:ENAB 258”
OUTPUT 724;”STAT:OPER:NTR?”
ENTER 724;N$
OUTPUT 724;”STAT:OPER:PTR?”
ENTER;P$
OUTPUT 724;”STAT:OPER:ENAB?”
ENTER 724;E$
Only one bit of the QUEStionable node is used:
• Bit 8 shows that there is an error in the calibration data.
:STATus:QUEStionable:NTRansition
Syntax
:STATus:QUEStionable:NTRansition <wsp> <value>
Содержание 8169A
Страница 36: ...Scanning the Poincare Sphere Analyzing the Results 38 Agilent 8169A Polarization Controller Second Edition ...
Страница 46: ...Programming the Polarization Controller GPIB Interface 48 Agilent 8169A Polarization Controller Second Edition ...
Страница 82: ...Remote Commands SYSTem Commands 84 Agilent 8169A Polarization Controller Second Edition ...
Страница 106: ...Installation 108 Agilent 8169A Polarization Controller Second Edition ...
Страница 114: ...Specifications 116 Agilent 8169A Polarization Controller Second Edition Declaration of Conformity ...
Страница 158: ...Cleaning Information Other Cleaning Hints 160 Agilent 8169A Polarization Controller Second Edition ...
Страница 166: ...Error Messages 168 Agilent 8169A Polarization Controller Second Edition ...
Страница 167: ......