Remote Commands
SRQ, The Service Request
60
Agilent
8169A
Polarization Controller, Second Edition
*RCL
Syntax
*RCL <wsp> <location>
0 _<location _<9
Definition
An instrument setting from the internal RAM is made the actual
instrument
setting (this does not include GPIB address or parser).
You recall user settings from locations 1-9. See \*SAV". Location 0
contains the default setting, which is the same as that obtained by *RST.
Example
OUTPUT 724;"*RCL 3"
*RST
Syntax
*RST
Definition
The reset setting (default setting) stored in ROM is made the actual
setting.
Instrument state: the instrument is placed in the idle state awaiting a
command.
The following are not changed:
• GPIB (interface) state
• Instrument interface address
• Output queue
• Service request enable register (SRE)
• Standard event status enable register (ESE)
The commands and parameters of the reset state are listed in
the following
table.
Содержание 8169A
Страница 36: ...Scanning the Poincare Sphere Analyzing the Results 38 Agilent 8169A Polarization Controller Second Edition ...
Страница 46: ...Programming the Polarization Controller GPIB Interface 48 Agilent 8169A Polarization Controller Second Edition ...
Страница 82: ...Remote Commands SYSTem Commands 84 Agilent 8169A Polarization Controller Second Edition ...
Страница 106: ...Installation 108 Agilent 8169A Polarization Controller Second Edition ...
Страница 114: ...Specifications 116 Agilent 8169A Polarization Controller Second Edition Declaration of Conformity ...
Страница 158: ...Cleaning Information Other Cleaning Hints 160 Agilent 8169A Polarization Controller Second Edition ...
Страница 166: ...Error Messages 168 Agilent 8169A Polarization Controller Second Edition ...
Страница 167: ......