Scanning the Poincare Sphere
Example: Measuring a Polarization Dependent Loss
36
Agilent
8169A
Polarization Controller, Second Edition
Set the Polarizing Filter
1
Press
on the polarization controller.
2
Press
3
Set the angle of the polarizing filter for maximum throughput.
a
Type in 10 and press
b
Press
twice to select the tens digit.
c
Using the Modify knob, increase the angle slowly until the power read on the
multimeter increases and then starts to decrease.
d
Press
once to select the units digit.
e
Using the Modify knob, decrease the angle slowly until the power read on the
multimeter starts to decrease.
f
Press
twice to select the hundredths digit
g
Using the Modify knob, increase the angle slowly until the power on the mul-
timeter starts to decrease. Return to the angle that gave the maximum pow-
er.
4
Connect the DUT into the setup, disturbing the setup as little as possible.
Figure 12 Setup with the DUT
Содержание 8169A
Страница 36: ...Scanning the Poincare Sphere Analyzing the Results 38 Agilent 8169A Polarization Controller Second Edition ...
Страница 46: ...Programming the Polarization Controller GPIB Interface 48 Agilent 8169A Polarization Controller Second Edition ...
Страница 82: ...Remote Commands SYSTem Commands 84 Agilent 8169A Polarization Controller Second Edition ...
Страница 106: ...Installation 108 Agilent 8169A Polarization Controller Second Edition ...
Страница 114: ...Specifications 116 Agilent 8169A Polarization Controller Second Edition Declaration of Conformity ...
Страница 158: ...Cleaning Information Other Cleaning Hints 160 Agilent 8169A Polarization Controller Second Edition ...
Страница 166: ...Error Messages 168 Agilent 8169A Polarization Controller Second Edition ...
Страница 167: ......