Remote Commands
SRQ, The Service Request
64
Agilent
8169A
Polarization Controller, Second Edition
So 16 would mean that the DSP (Digital Signal Processor)
Communications had failed, 18 would mean that the DSP Communications
had failed, and so had the Battery RAM. A value of zero shows no errors.
No further commands are allowed while the test is running.
The instrument is returned to the setting that was active at the time the
self-test query was processed.
The self-test does not require operator interaction beyond sending the
*TST? query.
Example
OUTPUT 724;"*TST?"
ENTER 724; A$
*WAI
Syntax
*WAI
Table 9 The Self Test Results
BITS
MNEMONICS
BIT VALUE
14
Motor 3
16384
13
Motor 2
8192
12
Motor 1
4096
10
Counter 3
1024
9
Counter 2
512
8
Counter 1
256
5
DSP Timeout
32
4
DSP Communications
16
3
Calibration Data
8
1
Battery RAM
2
0
Calibration Data Checksum
1
Содержание 8169A
Страница 36: ...Scanning the Poincare Sphere Analyzing the Results 38 Agilent 8169A Polarization Controller Second Edition ...
Страница 46: ...Programming the Polarization Controller GPIB Interface 48 Agilent 8169A Polarization Controller Second Edition ...
Страница 82: ...Remote Commands SYSTem Commands 84 Agilent 8169A Polarization Controller Second Edition ...
Страница 106: ...Installation 108 Agilent 8169A Polarization Controller Second Edition ...
Страница 114: ...Specifications 116 Agilent 8169A Polarization Controller Second Edition Declaration of Conformity ...
Страница 158: ...Cleaning Information Other Cleaning Hints 160 Agilent 8169A Polarization Controller Second Edition ...
Страница 166: ...Error Messages 168 Agilent 8169A Polarization Controller Second Edition ...
Страница 167: ......