Installation
Agilent
8169A
Polarization Controller, Second Edition
103
Humidity
The operating humidity for the Agilent 8169A is 15% to 95% from 0°C to
40°C.
Altitude
The Agilent 8169A may operate at up to 10,000 feet.
Installation Category and Pollution Degree
The Agilent 8169A has Installation Category II and Pollution Degree 2
according to IEC 60664.
Instrument Positioning and Cooling
Mount or position the instrument upright and horizontally so that air can
circulate around it freely. When operating the polarization controller,
choose a location that provides at least 75mm (3inches) of clearance at
the rear, and at least 25mm (1inch) of clearance at each side. Failure to
provide adequate air clearance may result in excessive internal temper
ature, reducing instrument reliability.
Figure 21 Correct Positioning of the Polarization Controller
Содержание 8169A
Страница 36: ...Scanning the Poincare Sphere Analyzing the Results 38 Agilent 8169A Polarization Controller Second Edition ...
Страница 46: ...Programming the Polarization Controller GPIB Interface 48 Agilent 8169A Polarization Controller Second Edition ...
Страница 82: ...Remote Commands SYSTem Commands 84 Agilent 8169A Polarization Controller Second Edition ...
Страница 106: ...Installation 108 Agilent 8169A Polarization Controller Second Edition ...
Страница 114: ...Specifications 116 Agilent 8169A Polarization Controller Second Edition Declaration of Conformity ...
Страница 158: ...Cleaning Information Other Cleaning Hints 160 Agilent 8169A Polarization Controller Second Edition ...
Страница 166: ...Error Messages 168 Agilent 8169A Polarization Controller Second Edition ...
Страница 167: ......