Extinction Ratio of Polarizer
Performance Test
Agilent
8169A
Polarization Controller, Second Edition
129
The Extinction Ratio is the absolute maximum of any measured value
(without the negative sign)
11
Note the measured result in your test record
12
Repeat list item 6 to list item 11 for 1510nm, 1530 nm and 1560 nm. Always
change wavelength setting on all instruments of the test setup, and make sure
to optimize the maximum transmission of the filter.
Example
The maximum measured value is one that might only occur momentarily
and after several tries because of random position changes, that is, you
need to approach the \minimum position" from both sides, for both
parameters several times.
N O T E
Table 25
Test
No.
Test Describtion
Minimum
Spec.
Result
Maximum
Spec.
Measurement
Certainty
3
Extinction Ratio
Wavelength
Extinction Ratio
1470 nm
40 dB
43.5 dB
1510 nm
45 dB
48.2 dB
1530 nm
45 dB
50.6 dB
1560 nm
40 dB
44.5 dB
Содержание 8169A
Страница 36: ...Scanning the Poincare Sphere Analyzing the Results 38 Agilent 8169A Polarization Controller Second Edition ...
Страница 46: ...Programming the Polarization Controller GPIB Interface 48 Agilent 8169A Polarization Controller Second Edition ...
Страница 82: ...Remote Commands SYSTem Commands 84 Agilent 8169A Polarization Controller Second Edition ...
Страница 106: ...Installation 108 Agilent 8169A Polarization Controller Second Edition ...
Страница 114: ...Specifications 116 Agilent 8169A Polarization Controller Second Edition Declaration of Conformity ...
Страница 158: ...Cleaning Information Other Cleaning Hints 160 Agilent 8169A Polarization Controller Second Edition ...
Страница 166: ...Error Messages 168 Agilent 8169A Polarization Controller Second Edition ...
Страница 167: ......