Setting the State of Polarization
Remote Commands
Agilent
8169A
Polarization Controller, Second Edition
69
[:INPut]:CIRCle:THETap
Syntax
[:INPut]:CIRCle:THETap <wsp> <value>jMINimumjMAXimumjDEFault
where value is a oating point number between -2160.00 and 2160.00.
Description
This command sets the position of the 2
θ
p
position on the Poincare
sphere. The parameter may be either
• a number, in optical degrees (do not give a unit; the number will be rounded to
the nearest 0.05°),
• MINimum (-2160.00°),
• MAXimum (2160.00°), or
• DEFault (0.00°).
[:INPut]:CIRCle:THETap?
Syntax
[:INPut]:CIRCle:THETap?
Description
This query gets the position of the 2
θ
p
position on the Poincare sphere in
optical degrees (without a unit).
The value you specify with this command is for 2
ε
B
.
N O T E
The value you specify with this command is for 2
θ
p
.
N O T E
The value
returned by this query
is for 2
θ
p
.
N O T E
Содержание 8169A
Страница 36: ...Scanning the Poincare Sphere Analyzing the Results 38 Agilent 8169A Polarization Controller Second Edition ...
Страница 46: ...Programming the Polarization Controller GPIB Interface 48 Agilent 8169A Polarization Controller Second Edition ...
Страница 82: ...Remote Commands SYSTem Commands 84 Agilent 8169A Polarization Controller Second Edition ...
Страница 106: ...Installation 108 Agilent 8169A Polarization Controller Second Edition ...
Страница 114: ...Specifications 116 Agilent 8169A Polarization Controller Second Edition Declaration of Conformity ...
Страница 158: ...Cleaning Information Other Cleaning Hints 160 Agilent 8169A Polarization Controller Second Edition ...
Страница 166: ...Error Messages 168 Agilent 8169A Polarization Controller Second Edition ...
Страница 167: ......