Oscilloscope Mode - The "Utilities" Menu
SCOPIX Oscilloscopes
IV - 87
Oscilloscope Mode (contd.)
I/O port config
RS 232
Selection of the "RS232" or "Network" menus for remote programming of the
"serial" "network" interface (ETHERNET).
This interface uses the connector (RS232 / ETHERNET) on the right-hand side
of the instrument. It requires the use of the RS232 / SUBD9 cable (HX0042).
Speed
Transmission speed:
300 to 115,200 Bauds.
Format
Word length:
7 or 8 bits.
Parity
Parity:
even, odd or no parity (none).
Stop bits
Number of stop bits:
1 or 2 stop bits
Protocol
Serial link management mode:
Hard
Hardware: the protocol is provided by the RTS and CTS lines
of the RS232 link.
Soft
Software: use of the XON and XOFF characters to synchronize
transmission and reception of the messages (reduced "3-wire" link)
None
No protocol checking
•
The «
»
symbol indicates the selected option.
•
The option can be modified using the stylus.
Network
Configuration of the ETHERNET parameters
provided by a DHCP
server
If this box is checked, the instrument makes a request to the network DHCP
server to automatically obtain:
•
an IP address
•
a subnet mask
•
the gateway address
•
possibly a network printer IP address
When powering up, if you activated this protocol during a previous
session, a confirmation is requested before sending the DHCP request
via ETHERNET.
If the instrument is no longer connected to the network, answer "no" to
the question, otherwise it will wait several minutes for an answer from the
DHCP server, before becoming functional.
Physical address
Oscilloscope address on the ETHERNET network.
This address cannot be modified (it is specific to the instrument)
Example:
00-01-02-03-04-63
IP address
Oscilloscope IP address on the ETHERNET network.
This address can be input automatically or manually with the keyboard, after
selecting the zone to be modified.
The
key can be used to delete the value preceding the cursor in the zone to
be modified.
An IP address can be assigned automatically by a DHCP server, if the server is
accessible, by ticking the box "provided by a DHCP server".
Example: 132.147.200.74
After modification, the IP address is displayed for 30s at the bottom right of the
screen.
Содержание OX7102-CII
Страница 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Страница 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Страница 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Страница 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Страница 59: ...Oscilloscope Mode The Trigger Menu SCOPIX Oscilloscopes IV 59 Oscilloscope Mode cont d The Trig Menu...
Страница 85: ...Oscilloscope Mode The Utilities Menu SCOPIX Oscilloscopes IV 85 Oscilloscope Mode cont d The Util Menu...
Страница 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Страница 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Страница 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Страница 143: ...Recorder Mode The Utilities Menu Recorder Mode cont d The Util Menu SCOPIX Oscilloscopes VII 143...
Страница 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Страница 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Страница 183: ...Applications SCOPIX Oscilloscopes IX 183...
Страница 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...