Oscilloscope Mode - The "Trigger" Menu
SCOPIX Oscilloscopes
IV - 69
Oscilloscope Mode (cont'd)
Triggered mode
Acquisitions and refreshment of the screen at each trigger event.
Automatic mode
Acquisition and automatic refreshing of screen even when there is no trigger
event.
Visible traces, even when there is no trigger event.
Single mode and
Roll mode
Acquisition of signal and refreshing of the screen on the first trigger occurring
after a trigger reset by pressing the key
opposite
(or via the time base menu).
In single-sweep mode, if the time base is over 50 ms/div, new samples are
displayed as soon as they are acquired and ROLL mode is activated as soon as
the acquisition memory is full (scrolling of the trace from the right to the left of the
screen).
The ROLL mode is not possible if the trigger on automatic measurement
threshold is active.
•
The " " symbol indicates the selected trigger mode.
•
The selected trigger mode is indicated in the status area
(Trig'd, Auto, Single).
•
The acquisition status is indicated in the status area: PRETRIG, RUN,
STOP, POSTRIG, READY, …
This selection can also be called up by double-pointing with the stylus on the
time base display area.
Triggering on
thresholds
Triggering on thresholds consists of:
1. acquisition implemented according to the conditions defined in the various
menus,
2. analyzing the trigger signal after classic acquisition,
3. searching for a condition on an automatic measurement,
4. if this condition is met, displaying the validated signals.
If "measurement on thresholds" is activated, the status zone at the foot of the
screen indicates it.
When threshold values are input, the instrument displays an error message if the
measurement of a threshold value is impossible.
Save / Restart
If this option is activated, all the traces acquired are saved to .TRC and .MAC
files after each acquisition.
As a result, you can save several rare events to the file system and analyze
them later.
The channels are saved in files, the name of which is determined according to
the date and time they are saved.
They can be saved on an FTP server or in the local file system if the FTP server
parameters have been configured (menu: "UTIL"
"Config I/O Ports"
"Network")
Содержание OX7102-CII
Страница 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Страница 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Страница 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Страница 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Страница 59: ...Oscilloscope Mode The Trigger Menu SCOPIX Oscilloscopes IV 59 Oscilloscope Mode cont d The Trig Menu...
Страница 85: ...Oscilloscope Mode The Utilities Menu SCOPIX Oscilloscopes IV 85 Oscilloscope Mode cont d The Util Menu...
Страница 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Страница 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Страница 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Страница 143: ...Recorder Mode The Utilities Menu Recorder Mode cont d The Util Menu SCOPIX Oscilloscopes VII 143...
Страница 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Страница 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Страница 183: ...Applications SCOPIX Oscilloscopes IX 183...
Страница 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...