Recorder Mode - The “Trigger” Menu
Recorder Mode (cont’d)
The "Trig" Menu
Triggering
Selection of trigger type and level on each channel. Triggering takes place if a
condition described by a line of the “Trigger” table is verified.
The trigger level should be defined in the channel measurement dynamic.
Source
Indicates the channel number.
Type
The type of trigger for each channel.
There are several possible types:
¾
None
¾
Lower than
¾
Higher than
¾
Lower/Higher
¾
Exterior
In "Recorder" mode, several conditions are monitored simultaneously.
Level 1
The main trigger threshold level can be set using the stylus.
Level 2
The auxiliary trigger threshold level can be set using the stylus. This tab is
enabled only if external trigger type is selected.
min duration
The fault will be evidenced if the fault condition defined by the type and
thresholds is present for a parameterizable period from 3,2 ms to 670 ks
according to the used time base (from 160 µs to 670 ks if the l’ EXTENDED
ACQUISITION MEMORY option).
*
Trigger levels are entered into the current value display area after modification.
Fine adjustment is possible.
SCOPIX Oscilloscopes
VII - 131
Содержание OX7102-CII
Страница 13: ...Description of the instrument SCOPIX Oscilloscopes II 13...
Страница 20: ...Description of the instrument II 20 SCOPIX Oscilloscopes...
Страница 21: ...Description of Accessories SCOPIX Oscilloscopes III 21...
Страница 34: ...Description of Accessories III 34 SCOPIX Oscilloscopes...
Страница 59: ...Oscilloscope Mode The Trigger Menu SCOPIX Oscilloscopes IV 59 Oscilloscope Mode cont d The Trig Menu...
Страница 85: ...Oscilloscope Mode The Utilities Menu SCOPIX Oscilloscopes IV 85 Oscilloscope Mode cont d The Util Menu...
Страница 93: ...Oscilloscope Mode SCOPIX Oscilloscopes IV 93...
Страница 108: ...Multimeter Mode V 108 SCOPIX Oscilloscopes...
Страница 118: ...Harmonic Analysis VI 118 SCOPIX Oscilloscopes...
Страница 143: ...Recorder Mode The Utilities Menu Recorder Mode cont d The Util Menu SCOPIX Oscilloscopes VII 143...
Страница 146: ...Recorder Mode VII 146 SCOPIX Oscilloscopes...
Страница 156: ...WEB Server VIII 156 SCOPIX Oscilloscopes...
Страница 183: ...Applications SCOPIX Oscilloscopes IX 183...
Страница 200: ...Technical Specifications X 200 SCOPIX Oscilloscopes...