6-24
ADS ECHO Manual
Problem
The monitor is recording “0” or negative depths
Possible
Causes
Values for manhole depth and/or physical offset in
QstartXML
are incorrect.
The temperature compensation values for March and/or
September are outside of reasonable values for the site.
Possible
Solutions
Verify manhole depth and physical offset values are
correct.
Ensure temperature compensation values for March and
September are reasonable.
Problem
Long Range Depth data is being measured correctly, but
UNIDEPTH
data is recording a straight line
Possible
Causes
A Tilt alarm is present
Possible
Solutions
Correct and reset the Tilt alarm