Configuration and Activation 4-3
Reactivating the Monitor
When the final configuration settings are complete, the user will need to reactivate the monitor. The
process will be the same as the
Activating the Monitor
section in
Chapter 2
, with the addition of setting the
Accelerometer and performing a test of the LRD sensor.
1.
Select the monitor that will be activated from the
Location Name
drop-down list in the
Monitor
section on the
QstartXML
main screen.
2.
Verify the configuration parameters and communication method (i.e. designated for
Connect
) for the
monitor and modify the parameters as necessary. If the
ECHO
modem is on (
COMMS
light is a slow
flashing green), the activation can be done wirelessly. If not, activate the
ECHO
using the Bluetooth
3.
Click on the
Activate
button in the
Functions
section.
QstartXML
initiates and establishes
communication with the monitor and then downloads the appropriate information and files. Once the
activation is complete, the status bar displays an
Activate Successful
message.
QstartXML
generates
an Activation log that is available for viewing through the log viewer within
QstartXML
.
4.
While still online with the
ECHO
, select the
Advanced
device. Under the
Diagnostics
tab, the
Accelerometer
must be set to establish a baseline against which it will compare later readings to check
for a
Tilt
alarm. Click the
Set
button to set the initial values. This will display reference values for the
X, Y, and Z. The
Read
button will display a comparison between current values and the initial base
values.
Setting the
Accelerometer
5.
Continue other activities that require communication with the monitor or disconnect the monitor by
clicking on the
Hangup
button in the
Functions
section. This will cause the local computer to
disconnect from the monitor.
6.
To ensure all settings have been configured properly, perform a test fire of the
Long Range Depth
sensor. If still online with the
ECHO
, open the
Long Range Depth
device dialog. Otherwise,
reconnect with the
ECHO
by selecting
Connect
, then open the
Long Range Depth
device dialog.
When online with the monitor, the dialog will change to allow the user to take readings. Click on the
Read
button to initiate the test. Observe the results to confirm proper sensor performance. The sensor
may need to be read several times for the readings to settle in after initial installation. If it appears that
the readings are too high and may be measuring the depth of the manhole bench rather than the depth
of flow in the invert, consider rotating the
Mounting Bar
slightly to provide the
ECHO
a more direct
line of sight to the invert. Get back online and repeat the
Long Range Depth
firing to verify that the
ECHO
is now reading correctly.
Содержание ECHO 9000
Страница 13: ...Introduction and Overview 1 9 First Page of the ECHO IECEx Certificate of Conformity ...
Страница 16: ...1 12 ADS ECHO Manual Sample ECHO Declaration of Conformity Page 1 ...
Страница 17: ...Introduction and Overview 1 13 Sample ECHO Declaration of Conformity Page 2 ...
Страница 61: ...Hardware Installation 3 21 Manhole Depth and Physical Offset Measurements ...