2-22
ADS ECHO Manual
Note:
For all future communications following the initial activation, change the
Connect
designation to
Wireless
to perform remote communications if the
ECHO
has a public static or
private static SIM card installed.
Verifying Wireless Communication
It is recommended that wireless communication be verified before installation of the monitor.
1.
Ensure an antenna is attached to the
ECHO
.
2.
Hold the magnet on the
Wake
activation point for 25 to 30 seconds until the
SENSE
LED illumines
magenta. Hold the magnet in place for another 1 to 2 seconds once this light illumines, then quickly
remove the magnet. The
SENSE
light will begin to rapidly flash a green light and the
COMMS
LED
will be a solid green until connected to the network. Once the green light is flashing, wireless
communications will be enabled.
3.
Change the connection type to
Wireless
and verify the IP address is correct.
4.
Verify the
COMMS
LED is a slow flashing green light.
5.
Attempt communication with the monitor by clicking the
Status
button.
6.
If the monitor status information is displayed, wireless communication has been verified.
Monitor Status
7.
Close the print preview of the status screen and click on the
Hangup
button in the
Functions
section
on the
QstartXML
main screen.
Saving the Configuration and Creating New Default
Location
1.
Save the configuration by clicking the
Save
button at the bottom of the
QstartXML
main screen.
2.
Clicking on the
Set as Default
button will update the default new location parameters to match how
the
ECHO
was just configured. This will allow the user to create a configuration template for all
subsequent
ECHO
monitor location. Clicking
New
will then populate the configuration with the
template parameters.
Содержание ECHO 9000
Страница 13: ...Introduction and Overview 1 9 First Page of the ECHO IECEx Certificate of Conformity ...
Страница 16: ...1 12 ADS ECHO Manual Sample ECHO Declaration of Conformity Page 1 ...
Страница 17: ...Introduction and Overview 1 13 Sample ECHO Declaration of Conformity Page 2 ...
Страница 61: ...Hardware Installation 3 21 Manhole Depth and Physical Offset Measurements ...