Only a pair of action values is recorded, including the switch state (on to off/off to on) when triggered and pressure
value.
Press
and select "Reset", to clear the action values.
Press
and select "Switch Setup", to switch between mechanical switch, NPN switch and PNP switch.
3.3.3 HART Communication
The calibrator supports HART bus communication, uses simplified DD files, and provides setting maintenance
and
calibration for general and common parameters of HART pressure transmitters. Before using the
calibrator to operate the
transmitter, please refer to the user manual of transmitter. If you need full-featured HART operation, please refer to Chapter
VII "HART communicator".
Note: During communication with a HART device, the calibrator acts as a master station all the time. To avoid damaging
the control system, you must separate the HART device from the control system before connecting the calibrator to HART
device.
1. Search and connection
Under the main operation interface, open the electrical measurement area for display. Select
in mode
switching to enable the HART function. The calibrator will automatically switch to power configuration selected previously
(connection mode of internal power supply and internal resistance by default), and search for the address "0". When a
HART device is found, the calibrator will automatically connect to it and display its reading.
Press
and select "Search", or click the HART measurement channel screen when no HART device is
Содержание 761A
Страница 1: ...761A Automated Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Страница 19: ......
Страница 20: ......
Страница 30: ...10 Warranty 1 year Optional Software ACal and Additel Land Additel Link for access via mobile application...
Страница 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Страница 33: ...13 LAN Interface USB port Power supply input Label Battery...
Страница 47: ...27 Figure 3 3 Gauge pressure outlet...
Страница 137: ...117 Figure 5 9 Calibration of the aneroid barometer Figure 5 10 Calibration of the barometric altimeter...
Страница 151: ...131...