107
the dwell time, and then click the OK button
to start the test. In case of the general digital barometer, the
readings need to be manually in put during the test. For the VAISALA digital barometer, the calibrator can
automatically obtain the readings through communication port.
Click the next button
to skip the current set point and enter the next set point for pressure control.
Selecting the manual execution requires manual input for each reading.
To stop this test, press the back button
.
4. Save the result
Click the save button
to save the test result.
The following are required fields: Name and serial number, click OK button
to complete the save.
Click the back button
to return to the test type list interface without saving.
Click the reset button
to rerun the test.
5.3.2 Mercury barometer
1. Prepare
Select “mercury barometer” in the test type list to enter to the quick test process of the mercury barometer.
2. Connection
Connect to the calibrator as shown in Figure 5-8.
3. Start the test
Содержание 761A
Страница 1: ...761A Automated Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Страница 19: ......
Страница 20: ......
Страница 30: ...10 Warranty 1 year Optional Software ACal and Additel Land Additel Link for access via mobile application...
Страница 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Страница 33: ...13 LAN Interface USB port Power supply input Label Battery...
Страница 47: ...27 Figure 3 3 Gauge pressure outlet...
Страница 137: ...117 Figure 5 9 Calibration of the aneroid barometer Figure 5 10 Calibration of the barometric altimeter...
Страница 151: ...131...