3. Task setting features
Select the task in the task list to enter the instrument parameter interface, click the start icon to enter the execution
information interface, and input the calibration information of the pressure gauge .
Table 5- 2 Task setting of analog pressure gauge
Item
Effective value
Description
Set point
s
The effective value of the set point
pressure depends on the pressure
range of the gauge; up to 17
calibration points can be set
Set the calibration points for the task, and the calibrator will
automatically set the default calibration point based on the range of
the calibrated instrument and the number of the calibration points. Any
point can be manually selected to be changed to a different point. The
set point needs to be within the range of the pressure gauge. Change
the number of the calibration points through the arrow key on the right
side of the screen or by clicking the number of the calibration points.
Cycle number
1, 2, 3
Select the number of cycles
Stroke
Round trip and one way
Select the calibration task stroke mode
The number of readings
Integer of 1 ~ 6
Input the number of readings to be taken when the calibrator reaches
the set point
Reading interval
Integer of (1 ~ 600) seconds
Interval time between two readings
Dwell time
Integer of (1 ~ 600) seconds
The waiting time before readings are taken and the calibrator
advances to the next set point.
Input format
Value, the number of scales
1. Value: Digital display of pressure value
2. The number of scales: The pressure value is displayed by counting
the number of scales to facilitate the reading and comparison with the
Содержание 761A
Страница 1: ...761A Automated Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Страница 19: ......
Страница 20: ......
Страница 30: ...10 Warranty 1 year Optional Software ACal and Additel Land Additel Link for access via mobile application...
Страница 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Страница 33: ...13 LAN Interface USB port Power supply input Label Battery...
Страница 47: ...27 Figure 3 3 Gauge pressure outlet...
Страница 137: ...117 Figure 5 9 Calibration of the aneroid barometer Figure 5 10 Calibration of the barometric altimeter...
Страница 151: ...131...