![Additel 761A Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/additel/761a/761a_user-manual_3047283068.webp)
48
Table 3- 5 auto step parameters
Item
Effective value
Description
Stroke
Round trip or one way
Set the trip pattern of auto step.
Cycle number
0-100
Set the cycle number of auto step.
Step mode
Step point, percentage, value,
or custom
Set the mode of automation step.
Cycle interval
0~3600 sec
Standing time between the end of each cycle and the start of next cycle.
Dwell time
0~3600 sec
Standing time after the present electrical output becomes stable.
Step point number
2-17
It can be set when the step mode is step point.
Range
Not to exceed the electrical
output range.
Set the output range of automatic step.
Step point list
Editable when the step mode
is custom; read-only and
displayed under other modes.
Display the set point list of automatic step.
Step value
6.25%" 00%
It can be set when the step mode is percentage.
Step value
1.5625~25mA
It can be set when the step mode is value.
6. Manual step
Press the
/
key on the keyboard or
on the screen, to manually step through values.
Содержание 761A
Страница 1: ...761A Automated Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Страница 19: ......
Страница 20: ......
Страница 30: ...10 Warranty 1 year Optional Software ACal and Additel Land Additel Link for access via mobile application...
Страница 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Страница 33: ...13 LAN Interface USB port Power supply input Label Battery...
Страница 47: ...27 Figure 3 3 Gauge pressure outlet...
Страница 137: ...117 Figure 5 9 Calibration of the aneroid barometer Figure 5 10 Calibration of the barometric altimeter...
Страница 151: ...131...