34
Table 3- 2 Scaling parameters
Item
Effective value
Description
Transfer function
Linear, square root, square
Type of the conversion function of the scaled
value
Input range
0%~100%
Percentage of the input range of scaling
Unit
Depend on the electrical measurement selected
Input unit of scaling
Output range
0%~100%
Percentage of the output range of scaling
Unit
User-editable field.
Output unit of scaling
Resolution
1, 0.1, 0.01, 0.001
Resolution of scaling
Filter: Provides a first order linear filter and a moving average filter. The moving average filter also allows setting
the extremum pair.
Table3- 3 Filtering parameters
Item
Effective value
Description
Filter type
First-order filter and average filter
Select the filtering mode.
Coefficient
0.01~1
Applicable to the first-order filter.
Filter sampling number
Integer 1~100
Sampling number of the average filter, correlated with
the sampling time.
Filter sampling time
0~20sec
Sampling time of the average filter, correlated with the
sampling number.
Содержание 761A
Страница 1: ...761A Automated Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Страница 19: ......
Страница 20: ......
Страница 30: ...10 Warranty 1 year Optional Software ACal and Additel Land Additel Link for access via mobile application...
Страница 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Страница 33: ...13 LAN Interface USB port Power supply input Label Battery...
Страница 47: ...27 Figure 3 3 Gauge pressure outlet...
Страница 137: ...117 Figure 5 9 Calibration of the aneroid barometer Figure 5 10 Calibration of the barometric altimeter...
Страница 151: ...131...