102
3. Task setting
Select the task in the task list to enter the instrument parameter interface, click the start icon to enter the execution information
interface, and input the calibration process information of the contact pressure gauge in turn:
Table 5- 10 Task setting of contact pressure gauge
Item
Effective value
Description
Set point list
Effective value of the set point
pressure depends on calibrated
gauge
Pressure range; up to 17 calibration
points can be set
Set the calibration point for the task, and the calibrator will automatically
set the default calibration points based on the range of the calibrated
instrument and the number of the calibration points. The value and
number of points can be adjusted as necessary.
Cycle number
1, 2, 3
Select the number of calibration process cycles
Stroke
Round trip and one way
Select the calibration task stroke mode
The number of readings
Integer of 1 ~ 6
Input the number of readings to be taken when the calibrator reaches
the set point
Reading interval
Integer of (1 ~ 600) seconds
Interval time between collection of two readings
Dwell time
Integer of (1 ~ 600) seconds
Waiting time from a stable set point the collection of all readings
Input format
Value, the number of scales
1. Value: Digital display of pressure value
2. The number of scales: The pressure value is displayed by counting
the number of scales to facilitate the reading and comparison with the
value indicated by the pressure gauge. The calibrated reading =
standard value + deviation number of scales X minimum scale division
Содержание 761A
Страница 1: ...761A Automated Pressure Calibrator...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Страница 19: ......
Страница 20: ......
Страница 30: ...10 Warranty 1 year Optional Software ACal and Additel Land Additel Link for access via mobile application...
Страница 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Страница 33: ...13 LAN Interface USB port Power supply input Label Battery...
Страница 47: ...27 Figure 3 3 Gauge pressure outlet...
Страница 137: ...117 Figure 5 9 Calibration of the aneroid barometer Figure 5 10 Calibration of the barometric altimeter...
Страница 151: ...131...