5.2.2 Digital pressure gauge
1. Connection
Connect the calibrator and digital pressure gauge as shown in Figure 5-1;
Figure 5- 1 Calibration of digital pressure
It is recommended to connect the reference port (REF) when calibrating a differential pressure gauge.
2. New task
Select the “digital pressure gauge” icon
in the test type list. The other contents are the same as those in Section
5.2.1 New Task for Calibration of the analog pressure gauge. The accuracy setting for calibration of the digital pressure
gauge is slightly different from the analog pressure gauge. The effective values of the accuracy for calibration of the digital
pressure gauge are 0.025%, 0.05%, 0.1%, 0.16%, 0.25%, 0.4%, 1 %, 1.6%, 2.5%, 4%, and custom.