53
Figure 3- 12
Calibration of a pressure gauge
4. When the pressure of each calibration point becomes stable and is displayed in green, record the readings of each point.
3.6.2 Pressure transmitter
The calibrator supports calibration of 2-wire, 3-wire and 4-wire pressure transmitters.
1. Prepare
Open the pressure output area of main operation interface: High-pressure or low-pressure range (depends on
the range of the transmitter).
Switch the electrical measurement item of main operation interface to: Current or voltage measurement (depend
on the output signal of the transmitter).