Guidance and manufacturer’s declaration –
electromagnetic immunity
The System is intended for use in the electromagnetic environment specified below. The customer or
the user of the System should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601
test level
Compliance
Level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines (100kHz
frequency)
± 1 kV for signal
lines (100kHz
frequency)
± 2 kV for
power supply
lines (100kHz
frequency)
± 1 kV for signal
lines (100kHz
frequency)
Mains power quality should be that
of a typical domestic, commercial,
or hospital environment.