Step
Action
5
(cont.)
You will see a butterfly on the screen while you
wait for the result. Do not remove the test strip
while the butterfly is on the screen. If sounds are
turned on, the Reader beeps once when the result
is ready.
If the butterfly does not appear, you may not have
applied enough control solution to the test strip.
Apply a second drop of control solution to the
test strip within 5 seconds of the first drop. If the
butterfly still does not appear or if more than 5
seconds have passed, discard the test strip. Turn
off the Reader and repeat the steps in this section
with a new test strip.
137