If you believe your glucose readings are not correct or are inconsistent
with how you feel, perform a blood glucose test on your finger to
confirm your glucose. If the problem continues, remove the current
Sensor, apply a new one, and contact Customer Service. Customer
Service is available at 1-855-632-8658 7 Days a Week from 8AM to 8PM
Eastern Standard Time.
What to know about the Reader’s Built-in Meter:
•
The FreeStyle Libre 2 Reader has a built-in blood glucose meter that is
designed to be used only with FreeStyle Precision Neo blood glucose
test strips and MediSense Glucose and Ketone Control Solution. Using
other test strips with the Reader’s built-in meter will produce an error
or cause the Reader’s built-in meter to not turn on or start a test. The
Reader’s built-in meter does not have ketone testing functionality.
•
The Reader’s built-in meter is not for use on people who are
dehydrated, hypotensive, in shock, or for individuals in hyperglycemic-
hyperosmolar state, with or without ketosis.
•
The Reader’s built-in meter is not for use on neonates, in critically-ill
patients, or for diagnosis or screening of diabetes.
•
See
Using the Reader’s Built-in meter
section for additional important
information on the use of the Reader’s built-in meter.
14