35
Electromagnetic Flowmeter
COPA-XE / MAG-XE
5.
Error Message
The following list of the error messages includes explanations for the Error Codes displayed
Error Codes 0 to 9, A, B, C do not occur during data entry.
Error Code
Detected System Error
Corrective Measures
0
Pipeline not full.
Open shut off devices; fill pipeline; adjust Empty Pipe Detector
1
2
3
A/D-Converter
Positive or negative reference too small.
Flowrate greater than 130 %.
Reduce flowrate, throttle shut off devices.
Check connection board and converter;
Reduce flowrate, change flow range
4
External zero return contact activated.
Zero Return activated by pump or field contact.
5
RAM defective
1. Error 5 appears in display
2. Error 5 appears only in Error Register
Program must be reinitiallized;
Contact ABB Automation Products Service department;
Information: Corrupted data in RAM, converter automatically executes a reset
and uploads data from EEPROM.
7
8
Positive reference too large
Negative reference too large.
Check signal cable and magnetic field excitation.
Check signal cable and magnetic field excitation.
6
9
A
B
C
Error > F
Error totalizer < R
Error totalizer
Excitation frequency defective
MAX-Alarm limit value
MIN-Alarm limit value
Primary data invalid
Reset forward totalizer or preset new values in totalizer.
Reset reverse totalizer or preset new values in totalizer.
Forward, reverse or difference totalizer defective.
Reset forward/reverse totalizer.
Check line frequency for Supply Power 50/60 Hz or for AC/DC Supply Power
Error in the Digital-Signal board.
Reduce flowrate
Increase flowrate
The data for the primary in the external EEPROM are invalid. Compare data in
submenu "Primary" with data listed on the Instrument Tag.
If the values agree, use "Store Primary" to reset the error message. If the data
are not identical, then it is necessary to first reenter the primary data and then
complete by using "Store Primary". Contact ABB Automation Products-Service
10
11
Entry >1.00 Q
maxDN
>10 m/s.
Entry <0.05 Q
maxDN
<0.5 m/s.
Reduce Q
max
range.
Increase Q
max
range.
16
17
Entry >10 % Low Flow Cutoff.
Entry < 0 % Low Flow Cutoff.
Reduce entry value.
Increase entry value.
20
21
22
Entry
≥
100 s Damping.
Entry <0.5 s Damping.
Entry >99 Instrument Address.
Reduce entry value.
Increase entry value, (as a function of the excitation frequency)
Reduce entry value.
38
39
40
41
Entry >1000 Pulses/Unit.
Entry < 0.001 Pulses/Unit.
Max. pulse frequency exceeded,
scaled pulse output.
Pulse factor (5 kHz).
Min. pulse frequency below limit
<0.00016 Hz.
Reduce entry value.
Increase entry value.
Reduce pulse factor.
Increase pulse factor
42
43
44
45
46
Entry >2000 ms Pulse Width.
Entry <0.1 ms Pulse Width.
Entry >5.0 g/cm3 Density.
Entry <0.01 g/cm3 Density.
Entry too large
Reduce entry value.
Increase entry value.
Reduce entry value.
Increase entry value.
Reduce pulse width entry value
54
56
Primary Zero > 50 Hz
Entry >3000 Threshold Empty Pipe
Detector.
Check ground and ground signals. Adjustment can be made if the flowmeter
primary is filled with fluid and the flowrate is zero.
Reduce entry value, check adjust "Empty Pipe Detector".
74/76
Entry > 130 % MAX - or MIN-Alarm
Reduce entry value
91
92
Data in EEPROM corrupted
Data in ext. EEPROM corrupted
Data in internal EEPROM invalid, for corrective measures see Error Code 5.
Data (e.g. Qmax, Damping) in external EEPROM invalid, access possible.
Occurs when function "Store data in ext. EEPROM" was not called. The error
message can be cleared using the function "Store data in ext. EEPROM"
93
94
Ext. EEPROM corrupted or not installed
Ext. EEPROM version incorrect
No access possible, component defective. If the component is not installed,
then it is necessary that the ext. EEPROM which belongs to the
flowmeter primary be installed.
The data base is not correct for the present software version. Calling the func-
tion "Load data from ext. EEPROM" initiates an automatic update of the exter-
nal data. The function "Store data in ext. EEPROM" clears the error message.
95
96
External Primary Data incorrect
EEPROM version incorrect
See Error Code C.
Data base in the EEPROM has a different version than the installed software.
Clear the error by calling the function "Update".
97
98
Primary incorrect
EEPROM corrupted or not installed
The flowmeter primary data in the internal EEPROM are invalid. The error can
be cleared by calling the function "Load Primary", (See Error Code C).
No access possible, component defective. If the component is not installed.
then it is necessary that the ext. EEPROM which belongs to the flowmeter
primary be installed.
99
99
Entry too large
Entry too small
Reduce entry value
Increase entry value