36
Electromagnetic Flowmeter
COPA-XE / MAG-XE
6.
Fuse Locations, Identification of the Converter Design,
ext. EEPROM Socket
The identification of the converter version can be found on the Instrument Tag, which is located on the metal frame
of the converter (see Illustration ).
Version 01
Current Pulse Output active
+ Contact Input
+ Contact Output
Version 02
Current Pulse Output active
+ Contact Input
+ Contact Output
+ HART Protocol
Version 03
Current Pulse Output passive
+ Contact Input
+ Contact Output
Version 04
Current Pulse Output passive
+ Contact Input
+ Contact Output
+ HART Protocol
Version 05
Current Pulse Output passive
+
--
+ Contact Output
+ RS 485
Version 06
--
+ Pulse Output passive
+
--
+ Contact Output
+ Profibus DP
Version 07
Current Pulse Output passive
+
--
+ Contact Output
+ Profibus PA
Fuse
A
Part No.
For low voltage
1A
D151B025U07
For high voltage
(85 - 253 V)
0.5 A
D151B025U06
Fuse V103
0.125 A
D151F003U14
Model No.:
Order No.:
Module No.:
U/fnom.:
Pmax:
Version:
E4000
9509N1187 / A324
D674A836U08
85 - 253 F AC, DC
< 14 VA
01
Supply Power
Socket
Ext. EEPROM
Fuse
Connection Plug
Magnet Coils
Fuse F103
Switch S202
Cert. Prot.
Pulse Output
passive
active
Jum
pers
Jum
pers
Display Board
Fig. 17
Converter Module COPA-XE/MAG-XE
Содержание COPA-XE
Страница 4: ......
Страница 24: ...18 Electromagnetic Flowmeter COPA XE MAG XE ...