Operating Instructions — Analyze and Process Data
System User Guide
4500 Series of Instruments
RUO-IDV-05-0264-A
87 of 138
Generate TWCs
A TWC shows total absorbance (mAU) on the y-axis plotted against time on the x-axis. For more
information about using the available icons, refer to
.
•
Click
Explore > Show > Show DAD TWC
.
The TWC opens in a pane below the DAD spectrum.
Adjust the Threshold
The threshold is an invisible line drawn parallel to the x-axis of a graph that sets a limit below
which the software will not include peaks in a spectrum. The line has a handle, represented by a
blue triangle to the left of the y-axis. Click the blue triangle to view a dotted line that represents
the threshold. The threshold can be raised or lowered, but changing the threshold value does not
change the data. The software does not label any peaks in the region that lies below the
threshold.
•
Do one of the following:
•
To raise the threshold, drag the blue triangle up the y-axis. To lower the
threshold, drag the blue triangle down.
•
Click
Explore > Set Threshold
. In the
Threshold Options
dialog that opens,
type the threshold value.
•
Click
Explore > Threshold
.
The graph updates to show the new threshold. Peak labeling and the peak list are
also updated.
Prerequisite Procedure
•
. Open a data file that contains a DAD spectrum.
Tip!
Right-click inside the pane containing the DAD spectrum and then
click
Show DAD TWC
.
Prerequisite Procedure
•
Tip!
To view the current threshold value, move the pointer over the
threshold handle.
Table 10-6 Right-Click Menu for Chromatogram Panes
Menu
Function
List Data
Lists the data points and integrates the peaks found in
chromatograms.
Show Spectrum
Generates a new pane.
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...