Operating Instructions — Acquisition Methods
4500 Series of Instruments
System User Guide
54 of 138
RUO-IDV-05-0264-A
Add a Period
•
In the
Acquisition method
pane, right-click the
Mass Spec
icon, and then click
Add
period
.
A period is added below the last period created.
Copy an Experiment into a Period
•
In the
Acquisition method
pane, press CTRL, and then drag the experiment to the
period.
The experiment is copied below the last experiment in the period.
Copy an Experiment within a Period
Use this procedure to add the same or similar experiments to a period if most or all of the
parameters are the same.
•
Right-click the experiment and then click
Copy this experiment
.
Scan Techniques
MS
: In MS scans, also referred to as single MS scans, ions are separated according to their
mass-to-charge (
m/z
) ratio. A single MS scan may be used to determine the molecular weight of
a compound. Single MS scans can also be referred to as survey scans. MS scans do not provide
any information as to the chemical make-up of the ions other than the
m/z
ratio. Perform MS/MS
or MS/MS/MS scan types to obtain more information about the ions.
MS/MS
: MS/MS scans are used to help identify or confirm a molecular species. In MS/MS scans,
a precursor ion can be fragmented in one of two locations.
•
For triple quadrupole systems, fragmentation occurs in the collision cell.
•
For LIT systems, fragmentation can occur in the collision cell or the linear ion trap.
If enough energy is used, the precursor ion fragments to produce characteristic product ions.
MS/MS/MS
: The LIT instrument MS/MS/MS scans go one step further than MS/MS scans. A
fragment that is produced in the collision cell is fragmented further in the trap to give more
structural information about the molecular ion.
Quadrupole-Mode Scan Types
Triple quadrupole instruments have high-sensitivity Multiple Reaction Monitoring (MRM)
capabilities required for quantitation experiments. In addition, they have highly specific scan
types, such as precursor ion and neutral loss scans, that allow a more advanced search to be
performed on the components of the samples.
Q1 MS (Q1)
: A full scan using the first quadrupole (Q1). The ion intensity is returned for every
mass in the scan range.
Prerequisite Procedures
• Open a multi-period method
Содержание QTRAP 4500 LC/MS/MS system
Страница 4: ...Revision Log 4500 Series of Instruments System User Guide 4 of 138 RUO IDV 05 0264 A ...
Страница 22: ...Hazards Symbols 4500 Series of Instruments System User Guide 22 of 138 RUO IDV 05 0264 A ...
Страница 36: ...Operating Instructions Hardware 4500 Series of Instruments System User Guide 36 of 138 RUO IDV 05 0264 A ...
Страница 76: ...Operating Instructions Batches 4500 Series of Instruments System User Guide 76 of 138 RUO IDV 05 0264 A ...
Страница 128: ...Service and Maintenance Information 4500 Series of Instruments System User Guide 128 of 138 RUO IDV 05 0264 A ...
Страница 130: ...Troubleshooting 4500 Series of Instruments System User Guide 130 of 138 RUO IDV 05 0264 A ...
Страница 136: ...4500 Series of Instruments System User Guide 136 of 138 RUO IDV 05 0264 A ...
Страница 138: ...Calibration Ions and Solutions 4500 Series of Instruments System User Guide 138 of 138 RUO IDV 05 0264 A ...